MXR-100/30
Product Specifications
Product Debion
The MXR-100/30 is an unipolar water-cooled tube assembly without integrated radiation protection. It comprises a single focal spot, a tungsten anode and a directional beam. The tube is specifically designed for thickness gauging and sorting applications.
Nominal tube voltage kV
Continuous rating 1000W
Focal spot acc. EN 12543 d = 5.5 mm
Focal spot acc. ASTM E1165-12 N/A
Filament current, max. 3.6 A
Filament voltage, typical V
Inherent filtration 0.8 ± 0.1 mm Be
Target material W
Target angle 30°
Radiation coverage 40° x 40°
Leakage radiation, max. at loading factors in 1m distance 2.5 mSv/h (100 kV; 10 mA)
Weight 2.1 kg
Terminal type CA11
Power rating charts
Power rating at given filament current
Connection of the filament
Cooling
Cooling medium Water
Cooling medium flow, min. 4 l/min
Cooling medium temperature at inlet, max. 35°; optimal 20°-35°, no water condensation /
kein Kondenswasser
Pressure at cooling medium inlet, max. 6 bar
Post-cooling time after switchoff, min. 2 min
Outline drawing
66.8
Die radiographischen Kenndaten sind dem Prüfschein des Strahlers zu entnehmen.
For radiographic characteristics of the tube refer to Final Test Report.
Masse ohne Toleranzangabe sind nur zur Inbation. Dimensions without tolerances are for inbation only.
General inbation
General
The control of the high voltage and the filament current as well as the design of the cooling system and the radiation protection are within the responsibility of the original equipment manufacturer (OEM). For detailed instructions refer to
our ?X-Ray Tube Manual“. Provisions in the generator must assure that the X-Ray tube will be protected against overcurrent, over- voltage and lack of cooling. Otherwise the tube and/or the radiation protection may be damaged and become a hazard.
Cooling
The customer is in charge that the cooling circuit is properly connected to the cooling system. Prior to operating the tube, the coolant must be turned on. Make sure that the coolant flow meets the required cooling conditions. When the tube is switched off, the coolant flow must continue for at least 2 minutes in order to protect the anode and the lead protection from overheating.
Radiation protection
When installing an X-ray tube assembly into a X-ray equipment and/or operating an X-ray equipment, the responsibility for radiation protection is with the user. Compliance with local regulatory requirements and limit valu- es must be assured. Do not modify the tube assembly itself!